Quick summary: O-Arm 1000 Imaging System - Base O-Arm. Date: 2023-08-04.
O-Arm 1000 Imaging System - Base O-Arm
2023-08-04
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: O-Arm 1000 Imaging System - Base O-Arm. Date: 2023-08-04.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
Medtronic identified O-Arm O2 imaging systems where an internal ground cable may be installed backward. When the cable is installed inverted, a portion of the uninsulated cable is not covered by the zippered jacket. Recall start date: July 25, 2023
What you should do
- Contact the manufacturer if you require additional information.