Public recall detail

Novum IQ Large Volume Pump

2022-09-23

medium
Recall Safety notice
Quick summary

Quick summary: Novum IQ Large Volume Pump. Date: 2022-09-23.

Recommended action

Recommended action: compare your product with the official notice and follow the instructions published by the official source.

Recall details

Quick summary: Novum IQ Large Volume Pump. Date: 2022-09-23.

Official details

Affected product details

Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.

BrandBaxter Healthcare Corporation
ProductNovum IQ Large Volume Pump
IssueMedical devices - Performance
Recall classType I
IdentificationRA-64580

Affected products

ProductNovum IQ Large Volume Pump
LotAll lots.
Model40700BAX
Serial numberAll lots.
DescriptionNovum IQ Large Volume Pump

Issue

Through ongoing internal testing, Baxter Corporation has observed isolated incident of over infusion on the Novum IQ LVP infusion pump during test runs at 0.1 ml/hr after 40 hours of continuous infusion within the 96-hour set change interval. Baxter Corporation is issuing a follow-up communication to notify users that all of the sets on the Novum IQ LVP compatible sets list are acceptable to use at rates below 1mL/hr except for the product code 1W5000, which will be removed from the Novum IQ LVP list of compatible sets. Recall start date: September 15, 2022

What you should do

  • Contact the manufacturer if you require additional information.
Additional official information
AudienceHealthcare
CategoryHealth products - Medical devices - General hospital and personal use
CompaniesBaxter Healthcare Corporation 1 Baxter Parkway, Deerfield, Illinois, United States, 60015
Published byHealth Canada
Recall classType I
Alert / recall typeHealth product recall
Identification numberRA-64580
Background{}
CompaniesBaxter Healthcare Corporation 1 Baxter Parkway, Deerfield, Illinois, United States, 60015
AudiencesHealthcare
Source updated2023-02-06