Public recall detail

Nova Statstrip® Ketone Test Strips

2026-11-05

medium
Health products Health product recall
Quick summary

Quick summary: Nova Statstrip® Ketone Test Strips. Reason: Health product recall. Category: Health products. Date: 2026-11-05.

Recommended action

Recommended action: compare your product with the official notice and follow the instructions published by the official source.

Recall details

Quick summary: Nova Statstrip® Ketone Test Strips. Reason: Health product recall. Category: Health products. Date: 2026-11-05.

Official details

Affected product details

Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.

BrandNova Biomedical Corporation
ProductNova Statstrip® Ketone Test Strips
IssueMedical devices - Performance
Recall classType II
IdentificationRA-82918

Affected products

ProductNova Statstrip® Ketone Test Strips
Lot2224295249
Model66841
Serial number2224295249
DescriptionNova Statstrip® Ketone Test Strips
ProductNova Statstrip® Ketone Test Strips
Lot2226117249
Model66841
Serial number2226117249
DescriptionNova Statstrip® Ketone Test Strips

Issue

Certain lots of StatStrip Ketone Test Strips were found to have an increased rate of quality control (QC) test failures during use. While patient whole-blood testing performance is not affected, the affected lots are being recalled to prevent delays in obtaining ketone test results. Users should immediately identify and discontinue their use, and follow the instructions provided in the recall notification.

What you should do

  • Contact the manufacturer if you require additional information.
Additional official information
AudienceHealthcare
CategoryHealth products - Medical devices - Chemistry
CompaniesNova Biomedical Corporation 200 Prospect St., Waltham, Massachusetts, United States, 02454
Recall date2026-10-23
Published byHealth Canada
Recall classType II
Alert / recall typeHealth product recall
Identification numberRA-82918
Original published date2026-11-05
Background{}
CompaniesNova Biomedical Corporation 200 Prospect St., Waltham, Massachusetts, United States, 02454
AudiencesHealthcare
Source updated2026-11-05

Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.