Quick summary: MOSAIQ. Reason: Performance. Category: General hospital and personal use. Date: 2026-05-06.
MOSAIQ
2026-05-06
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: MOSAIQ. Reason: Performance. Category: General hospital and personal use. Date: 2026-05-06.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
When using Anatomical Position Monitoring (APM) during a Comprehensive Motion Management (CMM) unity treatment delivery, if the user performs a baseline shift (BLS) plan and an error occurs during the plan transfer, in some instances, the BLS plan may not load into mosaiq resulting in the original beams being delivered without the new BLS applied.
What you should do
- Email: application.support.na@elekta.com
- Tel: +1 855 6935358
- https://www.elekta.com