Public recall detail

Monnal™ T60 Advanced

2026-04-30

medium
Anaesthesiology Performance
Quick summary

Quick summary: Monnal™ T60 Advanced. Reason: Performance. Category: Anaesthesiology. Date: 2026-04-30.

Recommended action

Recommended action: compare your product with the official notice and follow the instructions published by the official source.

Recall details

Quick summary: Monnal™ T60 Advanced. Reason: Performance. Category: Anaesthesiology. Date: 2026-04-30.

Official details

Affected product details

Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.

BrandAir Liquide Medical Systems
ProductMonnal™ T60 Advanced
IssueMedical devices - Performance
Recall classType II
IdentificationRA-81994

Affected products

ProductMonnal™ T60 Advanced
LotMT6A-02322
ModelKA017124
Serial numberMT6A-02322
DescriptionMonnal™ T60 Advanced
ProductMonnal™ T60 Advanced
LotMT6A-02324
ModelKA017124
Serial numberMT6A-02324
DescriptionMonnal™ T60 Advanced
ProductMonnal™ T60 Advanced
LotMT6A-02323
ModelKA017124
Serial numberMT6A-02323
DescriptionMonnal™ T60 Advanced
ProductMonnal™ T60 Advanced
LotMT6A-02325
ModelKA017124
Serial numberMT6A-02325
DescriptionMonnal™ T60 Advanced

Issue

Air Liquide Medical Systems has identified a manufacturing non-conformity that affects a limited and specific number of Monnal T60 and Monnal T60 advanced devices. It concerns the failure to fit washers beneath the screw head that secures the inspiratory unit of the device. The absence of a washer calls into question the conformity of the Monnal T60 units concerned with the requirements of standard EM 13718-1 (Requirements for medical devices used in air ambulances) and standard EN 1789+A1 (Requirements for medical devices used in road ambulances).

What you should do

  • Contact the manufacturer if you require additional information.
Additional official information
AudienceIndustry
CategoryHealth products - Medical devices - Anaesthesiology
CompaniesAir Liquide Medical Systems 6 Rue Georges Besse, Antony, France, 92182
Recall date2026-04-17
Published byHealth Canada
Recall classType II
Alert / recall typeHealth product recall
Identification numberRA-81994
Background{}
CompaniesAir Liquide Medical Systems 6 Rue Georges Besse, Antony, France, 92182
AudiencesIndustry
Source updated2026-04-30

Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.