Quick summary: Monnal T60. Date: 2023-03-24.
Monnal T60
2023-03-24
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Monnal T60. Date: 2023-03-24.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
During a ventilation sequence, the user started the "100% O2" function. This function lasts for two minutes but can be interrupted by pressing the 100% O2 key. Note: after the 100% O2 function has been used, the FiO2 setpoint originally entered is reapplied. The user wanted to change the FiO2 setpoint a few seconds before the end of the two-minute period of the 100% O2 function, and the confirmation of this setpoint took place just after the end of the 100% O2 function, which can cause a conflict between the setpoint applied and its display. Recall start date: March 13, 2023
What you should do
- Contact the manufacturer if you require additional information.
Additional official information
Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.
Frequently asked questions
Answers based on the official recall notice.
Product images
Official product images from Health Canada.