Quick summary: MOLLI 2 Systems. Date: 2026-01-23.
MOLLI 2 Systems
2026-01-23
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: MOLLI 2 Systems. Date: 2026-01-23.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
Stryker has updated the MOLLI 2 system Instructions for Use (IFU) following two medical device reports (MDRs) submitted to the FDA (June 5 2025). These reports involved patient injuries associated with MOLLI Marker dislodgement or dislocation due to magnetic attraction from magnetized surgical tools. The updated IFU now includes an additional warning that using magnetized surgical tools in close proximity to the marker during localization may lead to marker dislodgement.
What you should do
- Contact the manufacturer if you require additional information.
Additional official information
Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.
Frequently asked questions