Quick summary: Mixing Bowl And Spatula, 20- Pack. Reason: Sterility. Category: Orthopaedics. Date: 2026-04-14.
Mixing Bowl And Spatula, 20- Pack
2026-04-14
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Mixing Bowl And Spatula, 20- Pack. Reason: Sterility. Category: Orthopaedics. Date: 2026-04-14.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
Zimmer Biomet is conducting a lot specific medical device recall removal for the mixing bowl and spatula. Ten complaints have been received identifying issues at the time of use related to the package seal, including incomplete seals, wrinkles in the seals or peeling seals. Seal integrity testing conducted internally showed potential sterility breach in 7.5% of the tested units that had visual seal nonconformities. The devices are distributed in 20-pack boxes. The defectiveness was discovered on November 19, 2025. Issue was identified through the complaints handling process.
What you should do
- Contact the manufacturer if you require additional information.