Quick summary: MiniMed 670 g, 780 g, 770 g and 630 g . Date: 2024-10-16.
MiniMed 670 g, 780 g, 770 g and 630 g
2024-10-16
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: MiniMed 670 g, 780 g, 770 g and 630 g . Date: 2024-10-16.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
In some instances, pumps that have been dropped, bumped, or experienced physical impact, may experience damage to internal electrical components, which may result in reduced battery life, causing the battery alerts to occur when less battery life remains than the user guide states. Recall start date: October 11, 2024
What you should do
- Contact the manufacturer if you require additional information.
Additional official information
Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.
Frequently asked questions
Answers based on the official recall notice.
Product images
Official product images from Health Canada.