Public recall detail

MiniMed 670 g, 780 g, 770 g and 630 g

2024-10-16

medium
Recall Safety notice
Quick summary

Quick summary: MiniMed 670 g, 780 g, 770 g and 630 g . Date: 2024-10-16.

Recommended action

Recommended action: compare your product with the official notice and follow the instructions published by the official source.

Recall details

Quick summary: MiniMed 670 g, 780 g, 770 g and 630 g . Date: 2024-10-16.

Official details

Affected product details

Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.

BrandMedtronic Minimed
ProductMiniMed 670 g, 780 g, 770 g and 630 g
IssueMedical devices - Performance
Recall classType I
IdentificationRA-76339

Affected products

ProductMinimed 670g
LotAll lots.
ModelMMT-1782K MMT-1782 MMT-1782KRP MMT-1782CRP
Serial numberAll lots.
DescriptionMinimed 670g
ProductMinimed 670g Kit
LotAll lots.
ModelMMT-1762KCN MMT-1762CRP
Serial numberAll lots.
DescriptionMinimed 670g Kit
ProductMinimed 780g Kit
LotAll lots.
ModelMMT-1895CN
Serial numberAll lots.
DescriptionMinimed 780g Kit
ProductMinimed 630g
LotAll lots.
ModelMMT-1714K MMT-1714KRP MMT-1714
Serial numberAll lots.
DescriptionMinimed 630g
ProductMinimed 770g
LotAll lots.
ModelMMT-1881 MMT-1891CN
Serial numberAll lots.
DescriptionMinimed 770g

Issue

In some instances, pumps that have been dropped, bumped, or experienced physical impact, may experience damage to internal electrical components, which may result in reduced battery life, causing the battery alerts to occur when less battery life remains than the user guide states. Recall start date: October 11, 2024

What you should do

  • Contact the manufacturer if you require additional information.
Additional official information
AudienceGeneral public
CategoryHealth products - Medical devices - General hospital and personal use
CompaniesMedtronic Minimed 18000 Devonshire St., Northridge, California, United States, 91325
Published byHealth Canada
Recall classType I
Alert / recall typeHealth product recall
Identification numberRA-76339
Background{}
CompaniesMedtronic Minimed 18000 Devonshire St., Northridge, California, United States, 91325
AudiencesGeneral public
Source updated2024-10-18

Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.

Starting...

Related recalls

Similar recalls and safety notices.

Loading related recalls...