Quick summary: Microdeep Depth Electrode. Date: 2025-08-08.
Microdeep Depth Electrode
2025-08-08
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Microdeep Depth Electrode. Date: 2025-08-08.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
A manufacturing defect has been identified in the caps used in certain Dixi Medical devices. Although the cap is screwed onto the guiding screw, it is possible for the electrodes to slip inside. this indicates that the clamping function provided by the screw may not be fully effective, potentially compromising the safe positioning of the device. Customers are requested to isolate the affected devices. Recall start date: July 30, 2025
What you should do
- Retrieve and isolate the affected products.
Additional official information
Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.
Frequently asked questions
Answers based on the official recall notice.
Product images
Official product images from Health Canada.
Starting...
Explore more
Related recalls
Similar recalls and safety notices.