Quick summary: Merlin™ 3650 Patient Care System (PCS). Reason: Health product recall. Category: Health products. Date: 2026-05-22.
Merlin™ 3650 Patient Care System (PCS)
2026-05-22
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Merlin™ 3650 Patient Care System (PCS). Reason: Health product recall. Category: Health products. Date: 2026-05-22.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
Abbott is reaching out to customers to inform about a Merlin™ PCS 3650 programmer software update that addresses a software behavior that may occur during Pacing Capture Threshold (PCT) test of an AVEIR™ Leadless Pacemaker (LP). If intermittent loss of telemetry occurs during PCT test, the programmer may not successfully communicate the command to terminate the test, potentially leaving the LP in sub-threshold pacing until telemetry is fully disconnected. This issue can only occur during a PCT test, which is performed during implant or in clinic follow-up. This issue was identified through a review of complaints data, which confirmed four (4) complaints associated with the issue globally.
What you should do
- Contact the manufacturer if you require additional information.
Additional official information
Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.
Frequently asked questions
Answers based on the official recall notice.
Product images
Official product images from Health Canada.
Starting...
Explore more
Related recalls
Similar recalls and safety notices.