Quick summary: Megadyne Mega Soft Patient Return Electrode. Date: 2023-06-16.
Megadyne Mega Soft Patient Return Electrode
2023-06-16
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Megadyne Mega Soft Patient Return Electrode. Date: 2023-06-16.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
Megadyne Medical Products, Inc. (“Megadyne”) has received reports of patient burns identified after surgical procedures in which Mega Soft pads were used. Megadyne is aware of 63 complaints of serious patient burns globally since April 2018. Megadyne has conducted a thorough investigation, and to this point have not identified any design or manufacturing defects, nor determined definitive root cause. However, Megadyne has determined that in some instances the Mega Soft pad Instructions for Use (IFU) were not being properly followed. Recall start date: June 6, 2023
What you should do
- Contact the manufacturer if you require additional information.
Additional official information
Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.
Frequently asked questions
Answers based on the official recall notice.
Product images
Official product images from Health Canada.
Starting...
Explore more
Related recalls
Similar recalls and safety notices.