Public recall detail

Megadyne Mega Soft Patient Return Electrode

2023-06-16

medium
Recall Safety notice
Quick summary

Quick summary: Megadyne Mega Soft Patient Return Electrode. Date: 2023-06-16.

Recommended action

Recommended action: compare your product with the official notice and follow the instructions published by the official source.

Recall details

Quick summary: Megadyne Mega Soft Patient Return Electrode. Date: 2023-06-16.

Official details

Affected product details

Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.

BrandMegadyne Medical Products Inc.
ProductMegadyne Mega Soft Patient Return Electrode
IssueMedical devices - Labelling and packaging
Recall classType II
IdentificationRA-73835

Affected products

ProductMega Soft Universal Patient Return Electrode
LotAll lots that are within shelf-life
Model0846 0840 0845
Serial numberAll lots that are within shelf-life
DescriptionMega Soft Universal Patient Return Electrode
ProductMega Soft Dual Universal Plus Patient Return Electrode
LotAll lots that are within shelf-life
Model0848
Serial numberAll lots that are within shelf-life
DescriptionMega Soft Dual Universal Plus Patient Return Electrode
ProductMega2000 Patient Return Electrode
LotAll lots that are within shelf-life
Model0800
Serial numberAll lots that are within shelf-life
DescriptionMega2000 Patient Return Electrode
ProductMega2000 Soft Patient Pad Return Electrode
LotAll lots that are within shelf-life
Model0835 0830
Serial numberAll lots that are within shelf-life
DescriptionMega2000 Soft Patient Pad Return Electrode
ProductMega Soft Universal Plus Patient Return Electrode
LotAll lots that are within shelf-life
Model0847
Serial numberAll lots that are within shelf-life
DescriptionMega Soft Universal Plus Patient Return Electrode

Issue

Megadyne Medical Products, Inc. (“Megadyne”) has received reports of patient burns identified after surgical procedures in which Mega Soft pads were used. Megadyne is aware of 63 complaints of serious patient burns globally since April 2018. Megadyne has conducted a thorough investigation, and to this point have not identified any design or manufacturing defects, nor determined definitive root cause. However, Megadyne has determined that in some instances the Mega Soft pad Instructions for Use (IFU) were not being properly followed. Recall start date: June 6, 2023

What you should do

  • Contact the manufacturer if you require additional information.
Additional official information
AudienceHealthcare
CategoryHealth products - Medical devices - General and plastic surgery
CompaniesMegadyne Medical Products Inc. 4545 Creek Road, Cincinnati, Ohio, United States, 45242
Published byHealth Canada
Recall classType II
Alert / recall typeHealth product recall
Identification numberRA-73835
Background{}
CompaniesMegadyne Medical Products Inc. 4545 Creek Road, Cincinnati, Ohio, United States, 45242
AudiencesHealthcare
Source updated2023-07-14

Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.

Starting...

Related recalls

Similar recalls and safety notices.

Loading related recalls...