Quick summary: MEDRAD® Centargo Injection System. Date: 2025-01-28.
MEDRAD® Centargo Injection System
2025-01-28
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: MEDRAD® Centargo Injection System. Date: 2025-01-28.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
Centargo patient line sensor is a component of the Centargo injection system which detects a patient line. An investigation into reported air injection concerns found that, under certain special conditions, the patient line sensor may occasionally have difficulty detecting the presence of the patient line. This intermittent detection issue, combined with user reliance on automated system functions (such as auto-prime) and possible workflow inconsistencies, may slightly increase the likelihood of small air injections (1-10 ml). However, the overall risk remains low. Recall start date: December 13, 2024
What you should do
- Customers will be notified with instructions to follow.
Additional official information
Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.