Public recall detail

Medline Industries Scissors and Forceps

2024-04-05

medium
Recall Safety notice
Quick summary

Quick summary: Medline Industries Scissors and Forceps. Date: 2024-04-05.

Recommended action

Recommended action: compare your product with the official notice and follow the instructions published by the official source.

Recall details

Quick summary: Medline Industries Scissors and Forceps. Date: 2024-04-05.

Official details

Affected product details

Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.

BrandMedline Industries, Lp
ProductMedline Industries Scissors and Forceps
IssueMedical devices - Sterility
Recall classType II
IdentificationRA-75350

Affected products

ProductSterile Scissors
LotMore than 10 numbers, contact manufacturer.
ModelDYND04000 DYND04006
Serial numberMore than 10 numbers, contact manufacturer.
DescriptionSterile Scissors
Product5 Inch Curved Mosquito Forceps
LotMore than 10 numbers, contact manufacturer.
ModelDYND04009 DYNJ04048 DYND04010
Serial numberMore than 10 numbers, contact manufacturer.
Description5 Inch Curved Mosquito Forceps
ProductAdson Forceps, Standard, 4.75 In
Lot2023021590, 2023111390
ModelDYNJ04052
Serial number2023021590, 2023111390
DescriptionAdson Forceps, Standard, 4.75 In

Issue

This recall is due to having a weak seal, which may result in a breach of sterility of the contents if the seal fails. The weak seal may not be detectable by all users. Because of this failure and a breach of sterility, this can lead to an increased risk of contamination and infection and should not be added to the sterile field. Recall Start Date: March 6, 2024

What you should do

  • Contact the manufacturer if you require additional information.
Additional official information
AudienceHealthcare
CategoryHealth products - Medical devices - General and plastic surgery
CompaniesMedline Industries, Lp Three Lakes Drive, Northfield, Illinois, United States, 60093
Published byHealth Canada
Recall classType II
Alert / recall typeHealth product recall
Identification numberRA-75350
Background{}
CompaniesMedline Industries, Lp Three Lakes Drive, Northfield, Illinois, United States, 60093
AudiencesHealthcare
Source updated2024-04-05

Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.