Public recall detail

Marquee Disposable Core Biopsy Instrument Kit

2024-10-09

medium
Recall Safety notice
Quick summary

Quick summary: Marquee Disposable Core Biopsy Instrument Kit. Date: 2024-10-09.

Recommended action

Recommended action: compare your product with the official notice and follow the instructions published by the official source.

Recall details

Quick summary: Marquee Disposable Core Biopsy Instrument Kit. Date: 2024-10-09.

Official details

Affected product details

Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.

BrandBard Peripheral Vascular, Inc.
ProductMarquee Disposable Core Biopsy Instrument Kit
IssueMedical devices - Performance
Recall classType II
IdentificationRA-76311

Affected products

ProductMarquee Disposable Core Biopsy Instrument
LotMore than 10 numbers, contact manufacturer.
ModelMQ1210 MQ1213
Serial numberMore than 10 numbers, contact manufacturer.
DescriptionMarquee Disposable Core Biopsy Instrument
ProductMarquee Disposable Core Biopsy Instrument Kit
LotMore than 10 numbers, contact manufacturer.
ModelMQK1210 MQK1213
Serial numberMore than 10 numbers, contact manufacturer.
DescriptionMarquee Disposable Core Biopsy Instrument Kit

Issue

Based on customer complaints, Bard Peripheral Vascular, Inc. has identified that specific lots of the Marquee™ Disposable Core Biopsy Instruments may have been potentially manufactured missing the backend notch retention feature on the needle that results in inadequate securing of the needle inside the device. Recall Start Date: September 30, 2024

What you should do

  • Contact the manufacturer if you require additional information.
Additional official information
AudienceHealthcare
CategoryHealth products - Medical devices - Gastroenterology and urology
CompaniesBard Peripheral Vascular, Inc. 1625 West 3rd Street, Tempe, Arizona, United States, 85281
Published byHealth Canada
Recall classType II
Alert / recall typeHealth product recall
Identification numberRA-76311
Background{}
CompaniesBard Peripheral Vascular, Inc. 1625 West 3rd Street, Tempe, Arizona, United States, 85281
AudiencesHealthcare
Source updated2024-10-09

Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.