Public recall detail

Magnets in certain Philips Respironics sleep therapy masks may cause serious injury to some individuals

2022-09-26

critical
Recall Safety notice
Quick summary

Quick summary: Magnets in certain Philips Respironics sleep therapy masks may cause serious injury to some individuals. Date: 2022-09-26.

Recommended action

Recommended action: stop using or consuming the product if you believe it matches the recall, review the official details, and follow the instructions from the official source.

Recall details

Quick summary: Magnets in certain Philips Respironics sleep therapy masks may cause serious injury to some individuals. Date: 2022-09-26.

Official details

Affected product details

Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.

Brand
ProductPhilips Respironics sleep therapy masks with magnets in the headgear clips or straps (Amara View; DreamWisp; DreamWear; Wisp and Wisp Youth; and Therapy Mask 3100 NC/SP)
IssueHealth products - New safety information
IdentificationRA-64582

Affected products

ProductAmara View Full Face Mask (magnets are circled in black)
ProductDreamWisp Nasal Mask (magnets are circled in black)
ProductDreamWear Full Face Mask (magnets are circled in black)
ProductWisp and Wisp Youth Nasal Mask (magnets are circled in black)
ProductTherapy Mask 3100 NC/SP (magnets are circled in black)

Issue

Certain Philips Respironics sleep therapy masks used with Continuous Positive Airway Pressure (CPAP) or Bi-Level Positive Airway Pressure (BiPAP) machines have magnets in their headgear clips or straps. These magnets could cause safety concerns for people who may be vulnerable to magnetic fields. The sleep therapy mask magnets are used to attach the headgear clips or straps to the mask. Their magnetic field can potentially affect their function or displace some metallic medical devices or objects (e.g., pacemakers, metallic stents, metal in the eye) which may result in serious injury. Philips Respironics is adding a contraindication and is updating product warnings and instructions for use (IFU) to help prevent injury to patients and those close to them, who may be vulnerable to magnetic fields. Philips Respironics has received international reports of potential magnet interference with patient medical devices. These reports include: pacemaker alarms and interference, arrhythmia (irregular heart beat), pacemaker failure leading to replacement, need for shunt adjustment, resetting of automatic implantable cardioverter defibrillator, seizures, defibrillator shutting off periodically and cognitive issues. To date, no incidents have been reported in Canada. Health Canada is advising patients that the affected masks should not be used by patients if they or a person in close proximity to them including, household members, caregivers, or bed partners have any of the implanted metallic medical devices or objects listed below inside their bodies. For patients who have other metallic medical devices or objects inside their bodies that are not included on the list below but that may be affected by magnets, the manufacturer warns that the affected masks should be kept at least 6 inches (approximately 15 cm) away from the device/object. Please note this issue is not related to the recall of Philips Respironics CPAP and BiLevel PAP machines and mechanical ventilators due to issues involving the sound abatement foam, which Health Canada first communicated about in June 2021 and provided an update to in July 2022.

What you should do

  • Pacemakers
  • Implantable cardioverter defibrillators (ICD)
  • Neurostimulators
  • Magnetic metallic implants/electrodes/valves placed in upper limbs, torso, or higher (neck and head)
  • CSF (cerebral spinal fluid) shunts (e.g., VP [ventriculo peritoneal] shunt)
  • Aneurysm clips
  • Embolic coils
  • Intracranial aneurysm intravascular flow disruption devices
  • Metallic cranial plates, screws, burr hole covers, and bone substitute devices
  • Metallic splinters in the eye
  • Ocular implants (e.g., glaucoma implants, retinal implants)
  • Certain contact lenses with metal
  • Implants to restore hearing or balance that have an implanted magnet (such as cochlear implants, implanted bone conduction hearing devices, and auditory brainstem implants)
  • Magnetic denture attachments
  • Metallic gastrointestinal clips
  • Metallic stents (e.g., aneurysm, coronary, tracheobronchial, biliary)
  • Implantable ports and pumps (e.g., insulin pumps)
  • Hypoglossal Nerve Stimulators
  • Devices labelled as MR (magnetic resonance) unsafe
  • Magnetic metallic implants not labelled for MR or not evaluated for safety in a magnetic field
Additional official information
AudienceIdentification number RA-64582
CategoryHealth products - Medical devices
CompaniesPublished by Health Canada
Published byHealth Canada
Alert / recall typePublic advisory
Identification numberRA-64582
Background{}
CompaniesPublished by Health Canada
AudiencesIdentification number RA-64582
Source updated2022-09-26

Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.