Quick summary: Madsen® AccuScreen® TEOAE/DPOAE/ABR Probe. Reason: Performance. Category: Ear, nose and throat. Date: 2026-03-04.
Madsen® AccuScreen® TEOAE/DPOAE/ABR Probe
2026-03-04
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Madsen® AccuScreen® TEOAE/DPOAE/ABR Probe. Reason: Performance. Category: Ear, nose and throat. Date: 2026-03-04.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
There is already an inherent risk of a false pass with the DPOAE testing protocol with AccuScreen® for a child with a hearing disorder and this cannot be eliminated. This inherent risk of false negative answers was increased when deviating from the AccuScreen® factory default DPAOE Protocol (DP-1) using DPOAE Probe 8-69-41100.
What you should do
- PATH MEDICAL GMBH : FSCA@PATHME.DE
Additional official information
Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.
Frequently asked questions
Answers based on the official recall notice.