Quick summary: Luminos dRF Max, MOBILETT Elara Max, YSIO Max and YSIO X.Pree. Date: 2024-07-05.
Luminos dRF Max, MOBILETT Elara Max, YSIO Max and YSIO X.Pree
2024-07-05
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Luminos dRF Max, MOBILETT Elara Max, YSIO Max and YSIO X.Pree. Date: 2024-07-05.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
Siemens Healthineers was informed by the detector manufacturer that magnetic connectors may have been supplied as spare parts for portable detectors. If magnetic connection units are installed, a permanent (static) magnetic field is generated that exists independently of the operation of the detector. This can affect the function of implantable electronic devices such as pacemakers or implantable defibrillators if a minimum distance of >5 cm from the implanted device is not maintained. Recall Start Date: June 18, 2024
What you should do
- Contact the manufacturer if you require additional information.
Additional official information
Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.
Frequently asked questions
Answers based on the official recall notice.
Product images
Official product images from Health Canada.