Quick summary: Luer Slip and Luer Lock Syringes . Date: 2024-05-09.
Luer Slip and Luer Lock Syringes
2024-05-09
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Luer Slip and Luer Lock Syringes . Date: 2024-05-09.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
It has been identified that testing documentation does not support that the syringes can successfully deliver accurate volumes across the full range of their claimed graduated capacity and may pose a risk to patient health. Syringes not being able to deliver the intended dose may cause inaccurate dosing leading to the need for additional treatment or other medication. Recall start date: April 29, 2024
What you should do
- Contact the manufacturer if you require additional information.
Additional official information
Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.
Frequently asked questions
Answers based on the official recall notice.
Product images
Official product images from Health Canada.
Starting...