Quick summary: Liaison Q.S.E.T. Device Plus. Date: 2024-12-04.
Liaison Q.S.E.T. Device Plus
2024-12-04
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Liaison Q.S.E.T. Device Plus. Date: 2024-12-04.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
An investigation determined that 0.14% of the Liaison® Q.S.E.T. device plus in the lots identified have loose clear caps. The loose clear caps may allow leakage of the buffer from the tube. Recall start date: November 7, 2024
What you should do
- Instructions provided by manufacturer on inspecting and disposing of any affected devices.
Additional official information