Public recall detail

Leksell Gamma Knives

2026-02-18

medium
Radiology Performance
Quick summary

Quick summary: Leksell Gamma Knives. Reason: Performance. Category: Radiology. Date: 2026-02-18.

Recommended action

Recommended action: compare your product with the official notice and follow the instructions published by the official source.

Recall details

Quick summary: Leksell Gamma Knives. Reason: Performance. Category: Radiology. Date: 2026-02-18.

Official details

Affected product details

Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.

BrandElekta Solutions Ab
ProductLeksell Gamma Knives
IssueMedical devices - Performance
Recall classType II
IdentificationRA-81619

Affected products

ProductLeksell Gamma Knife® Perfexion™
LotAll lots.
Model715000
Serial numberAll lots.
DescriptionLeksell Gamma Knife® Perfexion™
ProductLeksell Gamma Knife® Icon™
LotAll lots.
Model1016200
Serial numberAll lots.
DescriptionLeksell Gamma Knife® Icon™
ProductLeksell Gamma Knife® Elekta Esprit™
LotAll lots.
Model1110200
Serial numberAll lots.
DescriptionLeksell Gamma Knife® Elekta Esprit™

Issue

When treating patients with the Leksell Gamma Knife®, the patient is attached to the device via an adapter. The adapter is the interface between the Leksell Gamma Knife® and the immobilization system used for the treatment. The Leksell Gamma Knife® has three types of adapters for patient fixation, the g-frame adapter, the vantage adapter, and the mask adapter. Elekta has become aware that a Quality Assurance (QA) procedure was initiated while the adapter was not properly locked to the patient positioning system of the Leksell Gamma Knife® as a result, the QA check failed.

What you should do

  • Elekta will issue a Field Safety Notice on 18 Feb 2026 for affected Gamme Knife and Elekta Esprit devices. The IFU will be updated to include a new warning.
Additional official information
AudienceHealthcare
CategoryHealth products - Medical devices - Radiology
CompaniesElekta Solutions Ab Hagaplan 4, Stockholm, Sweden, 11368
Recall date2026-02-18
Published byHealth Canada
Recall classType II
Alert / recall typeHealth product recall
Identification numberRA-81619
Background{}
CompaniesElekta Solutions Ab Hagaplan 4, Stockholm, Sweden, 11368
AudiencesHealthcare
Source updated2026-02-18

Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.

Starting...

Related recalls

Similar recalls and safety notices.

Loading related recalls...