Quick summary: iRICELL, IQ 200 and DxU Urine Analyzers. Date: 2024-02-07.
iRICELL, IQ 200 and DxU Urine Analyzers
2024-02-07
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: iRICELL, IQ 200 and DxU Urine Analyzers. Date: 2024-02-07.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
Beckman Coulter has become aware that the cannula, which is found in the flowcell and the IQclear rinse waste well assembly, may rust. Rust from the cannula has the potential to generate erroneous results. Recall start date: December 18, 2023
What you should do
- Contact the manufacturer if you require additional information.
Additional official information
Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.
Frequently asked questions
Answers based on the official recall notice.
Product images
Official product images from Health Canada.