Quick summary: iQ200 Series Analyzers and DxU 840m/850m Iris Analyzers. Reason: Health product recall. Category: Health products. Date: 2026-08-21.
iQ200 Series Analyzers and DxU 840m/850m Iris Analyzers
2026-08-21
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: iQ200 Series Analyzers and DxU 840m/850m Iris Analyzers. Reason: Health product recall. Category: Health products. Date: 2026-08-21.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
Beckman Coulter received customer complaints of lamina fluid leaks in iQ200 Series and DxU 840m/850m Analyzers. The investigation found that a structurally compromised rinse-line check valve may allow fluid to migrate internally and leak onto the table or floor, potentially resulting into exposure to physical or mechanical or chemical hazards.
What you should do
- Contact the manufacturer if you require additional information.
Additional official information
Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.
Frequently asked questions
Answers based on the official recall notice.
Product images
Official product images from Health Canada.
Starting...
Explore more
Related recalls
Similar recalls and safety notices.