Quick summary: Ionostar Plus and Ionolux. Date: 2024-02-02.
Ionostar Plus and Ionolux
2024-02-02
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Ionostar Plus and Ionolux. Date: 2024-02-02.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
Due to an error in the manufacturing process, it is possible that an incorrect plunger has been fitted to some capsules. This plunger does not contain the correct reactive component for Ionostar Plus. When these capsules are activated, curing is significantly delayed and incomplete. Processing according to the Instructions For Use (IFU) is not possible. The reactive component in the incorrect plungers contains small amounts of BHT and Methacrylate. These ingredients are not listed in the IFU. Recall Start Date: January 23, 2024
What you should do
- Contact the manufacturer if you require additional information.
Additional official information
Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.
Frequently asked questions
Answers based on the official recall notice.
Product images
Official product images from Health Canada.
Starting...
Explore more
Related recalls
Similar recalls and safety notices.