Public recall detail

Ionostar Plus and Ionolux

2024-02-02

medium
Recall Safety notice
Quick summary

Quick summary: Ionostar Plus and Ionolux. Date: 2024-02-02.

Recommended action

Recommended action: compare your product with the official notice and follow the instructions published by the official source.

Recall details

Quick summary: Ionostar Plus and Ionolux. Date: 2024-02-02.

Official details

Affected product details

Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.

BrandVoco Gmbh
ProductIonostar Plus and Ionolux
IssueMedical devices - Labelling and packaging
Recall classType II
IdentificationRA-75003

Affected products

ProductIonostar Plus
LotMore than 10 numbers, contact manufacturer.
Model2547 2543 2544 2549 2545 2548 2546
Serial numberMore than 10 numbers, contact manufacturer.
DescriptionIonostar Plus
ProductIonolux
LotMore than 10 numbers, contact manufacturer.
Model2119 2118 2117 2120 2121 2124 2122 2123
Serial numberMore than 10 numbers, contact manufacturer.
DescriptionIonolux
ProductIonolux
LotMore than 10 numbers, contact manufacturer.
Model2115
Serial numberMore than 10 numbers, contact manufacturer.
DescriptionIonolux
ProductIonostar Plus
LotMore than 10 numbers, contact manufacturer.
Model2540
Serial numberMore than 10 numbers, contact manufacturer.
DescriptionIonostar Plus

Issue

Due to an error in the manufacturing process, it is possible that an incorrect plunger has been fitted to some capsules. This plunger does not contain the correct reactive component for Ionostar Plus. When these capsules are activated, curing is significantly delayed and incomplete. Processing according to the Instructions For Use (IFU) is not possible. The reactive component in the incorrect plungers contains small amounts of BHT and Methacrylate. These ingredients are not listed in the IFU. Recall Start Date: January 23, 2024

What you should do

  • Contact the manufacturer if you require additional information.
Additional official information
AudienceHealthcare
CategoryHealth products - Medical devices - Dental
CompaniesVoco Gmbh Anton-Flettner, Cuxhaven, Germany, 27472
Published byHealth Canada
Recall classType II
Alert / recall typeHealth product recall
Identification numberRA-75003
Background{}
CompaniesVoco Gmbh Anton-Flettner, Cuxhaven, Germany, 27472
AudiencesHealthcare
Source updated2024-02-02

Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.