Quick summary: IOMAX Cortical Module. Date: 2024-05-24.
IOMAX Cortical Module
2024-05-24
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: IOMAX Cortical Module. Date: 2024-05-24.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
A manufacturing defect on some printed circuit board assemblies (PCBAs) within a single lot of PCBAs may cause damage to other internal components during assembly which may result in unintended electrical stimulation when using the transcranial stimulator and low current stimulator during the same procedure. IOMAX cortical modules manufactured between 02/15/2024 and 05/07/2024 may have been shipped with the defective printed circuit board. Recall start date: May 17, 2024
What you should do
- Contact the manufacturer if you require additional information.