Quick summary: IOMAX Cortical Module. Date: 2024-05-07.
IOMAX Cortical Module
2024-05-07
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: IOMAX Cortical Module. Date: 2024-05-07.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
An assembly error was discovered on the cortical module (PN 190296-200). A cortical module insulator that is used to provide an electric barrier between circuit boards for TCS-9, ICS, and other internal parts was found to be installed incorrectly. The insulator provides clearance between safety critical components and incorrect installation does not meet the minimum creepage/air gap required to meet IEC 60601-1. Recall start date: April 23, 2024
What you should do
- Contact the manufacturer if you require additional information. Return affected devices to manufacturer for repair or replacement.