Quick summary: Intramedullary Bone Saw Blade Assembly. Reason: Health product recall. Category: Health products. Date: 2026-07-03.
Intramedullary Bone Saw Blade Assembly
2026-07-03
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Intramedullary Bone Saw Blade Assembly. Reason: Health product recall. Category: Health products. Date: 2026-07-03.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
Zimmer Biomet is conducting a medical device recall removal for various sizes of the Intramedullary Bone Sawblade Assembly due to potential sterile barrier packaging damaged caused by the exposed saw blade. It was identified internally that the tip protector was incorrectly placed on the blunt end of the device instead of the sawblade, leaving the saw blade exposed. The exposed saw blade has potential to damage the sterile barrier pouch, which has potential to result in loss of sterility. There have been zero complaints received.
What you should do
- Contact the manufacturer if you require additional information.
Additional official information
Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.
Frequently asked questions
Answers based on the official recall notice.
Product images
Official product images from Health Canada.
Starting...
Explore more
Related recalls
Similar recalls and safety notices.