Quick summary: Ingenia and Achieva Systems. Reason: Performance. Category: Radiology. Date: 2026-04-16.
Ingenia and Achieva Systems
2026-04-16
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Ingenia and Achieva Systems. Reason: Performance. Category: Radiology. Date: 2026-04-16.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
Philips has become aware through the MRE software system developer Resoundant about an issue affecting MRE stiffness measurements when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm. Specifically, the reconstruction voxel size settings in the default MRE scan protocol are too small, which can potentially lead to inaccurate liver stiffness measurements (lower values).
What you should do
- Contact the manufacturer if you require additional information.