Quick summary: Ingenia 3.0T. Date: 2024-09-13.
Ingenia 3.0T
2024-09-13
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Ingenia 3.0T. Date: 2024-09-13.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
Philips has identified an issue in the MR system patient support where the integrated radio frequency (IRF) carrier assembly screws may come loose and protrude beyond the surface. This could cause the screws to interfere with other moving parts, potentially blocking the horizontal tabletop movement. The IRF carrier, a component of the patient support within the MR system, supports the tabletop where the patient lies. As the patient moves into the MRI tunnel, the tabletop, supported by the IRF carrier, forms a container for positioning the posterior coil. Recall start date: September 4, 2024
What you should do
- Contact the manufacturer if you require additional information.
Additional official information
Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.
Frequently asked questions
Answers based on the official recall notice.