Immulite 2000 System-Human Chorionic Gonadotropin (HCG) Assay
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Immulite 2000 System-Human Chorionic Gonadotropin (HCG) Assay.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
Siemens Healthcare Diagnostics Inc. has confirmed the potential for falsely elevated hcg results due to sample carryover. This can be observed when a sample is assayed for HCG immediately after an undiluted sample with a HCG value of GT;5000 MIU/mL. This issue impacts serum and urine patient samples, as well as quality control samples and adjustors. The effect of carryover varies based on the concentration of the high HCG sample. Recall start date: Mar 1, 2022