Icelock Ratchet Kit

2023-05-12

medium
Recall Safety notice
Quick summary

Quick summary: Icelock Ratchet Kit. Date: 2023-05-12.

Recommended action

Recommended action: compare your product with the official notice and follow the instructions published by the official source.

Recall details

Quick summary: Icelock Ratchet Kit. Date: 2023-05-12.

Official details

Affected product details

Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.

BrandOssur Canada Inc.
ProductIcelock Ratchet Kit
IssueMedical devices - Performance
Recall classType II
IdentificationRA-73641

Affected products

ProductIcelock 621 Ratchet Adaption Kit [L-621100]
LotContact the manufacturer.
ModelNot applicable.
Serial numberContact the manufacturer.
DescriptionIcelock 621 Ratchet Adaption Kit [L-621100]
ProductIcelock 621 Ratchet Adaption Kit [L-621100]
LotContact the manufacturer.
ModelNot applicable.
Serial numberContact the manufacturer.
DescriptionIcelock 621 Ratchet Adaption Kit [L-621100]
ProductIcelock 125 Ratched [ L-125000]
LotContact the manufacturer.
ModelNot applicable.
Serial numberContact the manufacturer.
DescriptionIcelock 125 Ratched [ L-125000]
ProductIcelock 600xm Ratche [L-621200]
LotContact the manufacturer.
ModelNot applicable.
Serial numberContact the manufacturer.
DescriptionIcelock 600xm Ratche [L-621200]
ProductIcelock 621 Ratchet [L-621000]
LotContact the manufacturer.
ModelNot applicable.
Serial numberContact the manufacturer.
DescriptionIcelock 621 Ratchet [L-621000]

Issue

The attachment pin and lock can wear down faster than expected in certain circumstances. Faster wear can result in the potential for the pin to disengage from the lock, leading to loss of suspension, potentially leading to an injury. Recall start date: April 24, 2023

What you should do

  • Contact the manufacturer if you require additional information. Notify, replace &/or return affected lots for both lock body & attachment pin as directed.
Additional official information
AudienceHealthcare Industry
CategoryHealth products - Medical devices - Physical medicine
CompaniesOssur Canada Inc. 2150-6900 Graybar Road, Richmond, British Columbia, Canada, V6W 0A5
Published byHealth Canada
Recall classType II
Alert / recall typeHealth product recall
Identification numberRA-73641
Background{}
CompaniesOssur Canada Inc. 2150-6900 Graybar Road, Richmond, British Columbia, Canada, V6W 0A5
AudiencesHealthcare Industry
Source updated2023-05-12