Quick summary: i‑STAT EG7+ Cartridge. Reason: Performance. Category: Chemistry. Date: 2026-04-15.
i‑STAT EG7+ Cartridge
2026-04-15
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: i‑STAT EG7+ Cartridge. Reason: Performance. Category: Chemistry. Date: 2026-04-15.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
Cartridges from the specific lots listed may report higher than expected pCO2 and lower than expected pH results due to a manufacturing issue which may present a pattern suggestive of severe respiratory acidosis. If such results are acted upon without consideration of the patient's overall clinical presentation, this could lead to unnecessary or inappropriate clinical interventions. Use of the cartridges should be discontinued for pH and pCO2.
What you should do
- Contact the manufacturer if you require additional information.