Quick summary: Hollow Fiber Hemodialyzers. Date: 2025-09-16.
Hollow Fiber Hemodialyzers
2025-09-16
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Hollow Fiber Hemodialyzers. Date: 2025-09-16.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
The dialyzer caps had been changed from a threaded version to a push on version whereby the user must push the caps firmly onto the dialyzer to securely affix the cap. As users were accustomed to twisting the caps on, there were complaints of caps coming off resulting in dialysate leaks. two (2) of the complaints received in the United States reported a serious injury to the user related to a slip/fall resulting from the fluid leak. As part of CAPA actions by the manufacturer, IFU (F50010927 rev. B) was updated to IFU (F50010927 rev. C) to specify the method for securing the blood port cap to the dialysate port. As part of this recall, the updated IFU instructions will be communicated.
What you should do
- Contact the manufacturer if you require additional information.
Additional official information
Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.
Frequently asked questions
Answers based on the official recall notice.
Product images
Official product images from Health Canada.
Starting...
Explore more
Related recalls
Similar recalls and safety notices.