Hls Set Advanced 5.0 and 7.0 With Bioline Coating
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Hls Set Advanced 5.0 and 7.0 With Bioline Coating.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
Maquet Cardiopulmonary Gmbh (MCP) has received customer complaints of damage to the primary packaging of the Hls Set Advanced. In all it was reported that the Tyvek cover was damaged. Investigation suggest that the Tyvek damage only occurs if the set drops upside down so either the velcro straps, holding the Hls modules, come loose or the Hls module detaches. A shift of the Hls module may cause the venous probe tab(s) to perforate the Tyvek cover. Possible sterility compromise. Recall start date: July 18, 2022.
What you should do
- Contact the manufacturer if you require additional information.