Quick summary: High Flow Insufflation Tubing (With Filter). Date: 2024-08-15.
High Flow Insufflation Tubing (With Filter)
2024-08-15
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: High Flow Insufflation Tubing (With Filter). Date: 2024-08-15.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
The unit package seal failed and it compromised sterility. Although one defect was detected in the field, no adverse events have occurred as a result of this incident. Inspection of inventory on-hand determined that there may be around 1% of this product (p/n 20400161S, lot #185198) with opened seals and if used could result in patient infections. Recall Start Date: July 19, 2024
What you should do
- Contact CANADIAN HOSPITAL SPECIALTIES LTD. (CHS) at RECALLS@CHSLTD.COM.
Additional official information
Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.
Frequently asked questions
Answers based on the official recall notice.
Product images
Official product images from Health Canada.
Starting...