Public recall detail

HeartMate II® LVAS - System Controller

2025-10-17

medium
Recall Safety notice
Quick summary

Quick summary: HeartMate II® LVAS - System Controller. Date: 2025-10-17.

Recommended action

Recommended action: compare your product with the official notice and follow the instructions published by the official source.

Recall details

Quick summary: HeartMate II® LVAS - System Controller. Date: 2025-10-17.

Official details

Affected product details

Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.

BrandTHORATEC CORPORATION
ProductHeartMate II® LVAS - System Controller
IssueMedical devices - Performance
Recall classType I
IdentificationRA-78314

Affected products

ProductHeartMate II® LVAS - System Controller
Lots/n: PCX-22221
Model106762
Serial numbers/n: PCX-22221
DescriptionHeartMate II® LVAS - System Controller
ProductHeartMate II® LVAS - System Controller
Lots/n: PCX-22321
Model106762
Serial numbers/n: PCX-22321
DescriptionHeartMate II® LVAS - System Controller
ProductHeartMate II® LVAS - System Controller
Lots/n: PCX-22270
Model106762
Serial numbers/n: PCX-22270
DescriptionHeartMate II® LVAS - System Controller
ProductHeartMate II® LVAS - System Controller
Lots/n: PCX-22271
Model106762
Serial numbers/n: PCX-22271
DescriptionHeartMate II® LVAS - System Controller

Issue

Abbott is initiating a voluntary medical device recall for the HeartMate II® and HeartMate 3™ System Controllers. This recall does not involve device removal from where they are used or sold. Abbott has observed an increase in customer complaints about the controller Backup Battery Fault Alarm. This alarm is indicated by a flashing yellow wrench on the user interface of the System Controllers. When the controller is connected to an external power source (e.g. 14V batteries or AC power via the Mobile Power Unit), this is an advisory alarm and does not indicate a critical system failure or impact pump functionality. To reduce future instances of the Backup Battery Fault Alarm, Abbott is advising that when performing backup battery installation or replacement, clinicians should minimize excessive handling, movement, stress and pulling at the interface between the ribbon cable and controller. An example of excessive handling would be using the ribbon cable to hold the weight of the controller. Please remind patients and caregivers to follow the patient handbook: if System Controller Backup Battery Fault Alarm is active, patients should first ensure that no other alarms are active, and then they should call their hospital contact as soon as possible for diagnosis and instructions. Patients and caregivers should not immediately attempt to exchange the system controller without clinical guidance.

What you should do

  • Contact the manufacturer if you require additional information.
Additional official information
AudienceHealthcare
CategoryHealth products - Medical devices - Cardiovascular
CompaniesThoratec Corporation 6035 Stoneridge Drive, Pleasanton, California, United States, 94588
Recall date2025-10-10
Published byHealth Canada
Recall classType I
Alert / recall typeHealth product recall
Identification numberRA-78314
Background{}
CompaniesThoratec Corporation 6035 Stoneridge Drive, Pleasanton, California, United States, 94588
AudiencesHealthcare
Source updated2025-10-17