Quick summary: Habib™ EndoHPB Catheter. Date: 2023-03-16.
Habib™ EndoHPB Catheter
2023-03-16
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Habib™ EndoHPB Catheter. Date: 2023-03-16.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
During investigational testing of the Habib™ Catheter with the ERBE VIO 3 generator (compatible 3rd party device), an issue with the generator setting in the Habib™ IFU (50738103) was identified. From the results of this testing, it was determined that the VIO 3 setting produces results that do not meet the dimensional requirements identified for the ablation heating zone in the Habib™ design matrix (92252607). Recall start date: March 2, 2023
What you should do
- Contact the manufacturer if you require additional information.
Additional official information
Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.
Frequently asked questions
Answers based on the official recall notice.
Product images
Official product images from Health Canada.