Quick summary: Giraffe™ OmniBed™ Carestation™. Reason: Health product recall. Category: Health products. Date: 2026-06-04.

Official details

Affected product details

Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.

BrandDatex-Ohmeda, Inc.
ProductGiraffe™ OmniBed™ Carestation™
IssueMedical devices - Performance
Recall classType II
IdentificationRA-82160

Affected products

ProductGiraffe™ OmniBed™ Carestation™
LotMore than 10 numbers, contact manufacturer.
Model2082844-001
Serial numberMore than 10 numbers, contact manufacturer.
DescriptionGiraffe™ OmniBed™ Carestation™
ProductGiraffe™ OmniBed™
LotMore than 10 numbers, contact manufacturer.
Model2074920-001-XXXXXX
Serial numberMore than 10 numbers, contact manufacturer.
DescriptionGiraffe™ OmniBed™

Issue

GE HealthCare has become aware of a potential issue involving the canopy soffit beneath the control panel end of the device. Under certain conditions, for example, impact during transport, one or more of the fasteners that secure the soffit can disengage. If this occurs, a fastener could fall into the patient compartment. In rare circumstances, if multiple fasteners disengage, the canopy soffit can become loose or detach. These conditions could potentially result in patient injury.

What you should do

  • Contact the manufacturer if you require additional information.
Additional official information
AudienceHealthcare
CategoryHealth products - Medical devices - General hospital and personal use
CompaniesDatex-Ohmeda, Inc. 3114 N Grandview Blvd, Waukesha, Wisconsin, United States, 53188
Recall date2026-05-25
Published byHealth Canada
Recall classType II
Alert / recall typeHealth product recall
Identification numberRA-82160
Background{}
CompaniesDatex-Ohmeda, Inc. 3114 N Grandview Blvd, Waukesha, Wisconsin, United States, 53188
AudiencesHealthcare
Source updated2026-06-04

Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.

Starting...

Related recalls

Similar recalls and safety notices.

Loading related recalls...