Quick summary: Getinge Washer/Disinfector. Date: 2025-08-28.
Getinge Washer/Disinfector
2025-08-28
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Getinge Washer/Disinfector. Date: 2025-08-28.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
The issue found can cause the device to overheat during operation if the circulation pressure calibration was not performed or been performed incorrect and the low-pressure alarm is not triggered. The scope of this action is the electrical heated 88-5 washer disinfectors. Justification: the construction on steam heated devices is different and the heating elements cannot be overheated. This field action is initiated to verify that the circulation pressure in standby mode reads 0.0 kPa. If the value deviates from 0.0 kPa, the device must be recalibrated by a qualified technician to ensure safe continued clinical use. Recall start date: August 21, 2025
What you should do
- Contact the manufacturer if you require additional information.
Additional official information
Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.
Frequently asked questions
Answers based on the official recall notice.