Quick summary: FreeO2 Automated Oxygen Therapy Device Adult & Pediatric - Version HW 2.3. Reason: Health product recall. Category: Health products. Date: 2026-10-08.
FreeO2 Automated Oxygen Therapy Device Adult & Pediatric - Version HW 2.3
2026-10-08
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: FreeO2 Automated Oxygen Therapy Device Adult & Pediatric - Version HW 2.3. Reason: Health product recall. Category: Health products. Date: 2026-10-08.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
Under specific conditions following a deep battery discharge, error 29515 may occur during startup of a FreeO2 HW 2.3 device using software package 1.4.2 and prevent its use. The update to software package 1.4.3 corrects this issue and restores the expected startup behaviour.
What you should do
- Contact the manufacturer if you require additional information.
Additional official information
Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.
Frequently asked questions
Answers based on the official recall notice.
Product images
Official product images from Health Canada.
Starting...
Explore more
Related recalls
Similar recalls and safety notices.