Quick summary: Flowpra Class I And II Screening Tests. Date: 2024-06-20.
Flowpra Class I And II Screening Tests
2024-06-20
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Flowpra Class I And II Screening Tests. Date: 2024-06-20.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
A suspected false negative for HLA class I with FL1-30 lot 022. Complaint investigation testing was performed using 4 samples with low levels of A2 antibody specificities and 1 sample with low levels of A31 antibody specificities. Investigation confirmed a lower percentage of positive reactions for samples tested with FL1-30 lot 022. In addition, investigation noted that lot specific user document, TDX-OLI-DMR-LOT-8536 revision 01, has incorrect reference values for the distribution of certain HLA class I antigen specificities. Recall start date: May 29, 2024
What you should do
- Instructions provided on response.
Additional official information
Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.
Frequently asked questions
Answers based on the official recall notice.
Product images
Official product images from Health Canada.
Starting...