Public recall detail

Flex. Biopsy and Grasping Forceps

2023-04-05

medium
Recall Safety notice
Quick summary

Quick summary: Flex. Biopsy and Grasping Forceps. Date: 2023-04-05.

Recommended action

Recommended action: compare your product with the official notice and follow the instructions published by the official source.

Recall details

Quick summary: Flex. Biopsy and Grasping Forceps. Date: 2023-04-05.

Official details

Affected product details

Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.

BrandRichard Wolf GmbH
ProductFlex. Biopsy and Grasping Forceps
IssueMedical devices - Performance
Recall classType II
IdentificationRA-73456

Affected products

ProductFlex. Grasping Forceps 5fr WL 550 Mm
LotMore than 10 numbers, contact manufacturer.
ModelNot applicable.
Serial numberMore than 10 numbers, contact manufacturer.
DescriptionFlex. Grasping Forceps 5fr WL 550 Mm
ProductFlex. Biopsy Forceps 5fr WL 550mm
LotMore than 10 numbers, contact manufacturer.
Model829.051
Serial numberMore than 10 numbers, contact manufacturer.
DescriptionFlex. Biopsy Forceps 5fr WL 550mm
ProductFlex. Grasping Forceps 5fr WL 550 Mm Ii
LotMore than 10 numbers, contact manufacturer.
ModelNot applicable.
Serial numberMore than 10 numbers, contact manufacturer.
DescriptionFlex. Grasping Forceps 5fr WL 550 Mm Ii
ProductFlex. Grasp. Forceps 6.6fr WL 550mm
LotMore than 10 numbers, contact manufacturer.
ModelNot applicable.
Serial numberMore than 10 numbers, contact manufacturer.
DescriptionFlex. Grasp. Forceps 6.6fr WL 550mm

Issue

All forceps can also be opened when the distal endoscope is fully angled and the proximal endoscope shaft is bent up to max. 90° and the proximal forceps area is not tilted in relation to the endoscope. However, when the endoscope shaft is bent by approx. 90°, the jaw opening is no longer fully possible, and the first restrictions on use begin from this positioning. If, in addition, the proximal forceps shaft is bent, the forceps can no longer be opened or restrictions begin even with a slight bending of the endoscope. Recall start date: March 9, 2023

What you should do

  • Contact AMT Surgical: INFO@AMTSURGICAL.COM.
Additional official information
AudienceHealthcare
CategoryHealth products - Medical devices - Gastroenterology and urology Health products - Medical devices - General and plastic surgery
CompaniesRichard Wolf GmbH Pforzheimer Str.32, Knittlingen, Baden-Wuerttemberg, Germany, 75438
Published byHealth Canada
Recall classType II
Alert / recall typeHealth product recall
Identification numberRA-73456
Background{}
CompaniesRichard Wolf GmbH Pforzheimer Str.32, Knittlingen, Baden-Wuerttemberg, Germany, 75438
AudiencesHealthcare
Source updated2023-04-05

Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.