Public recall detail

Flamingo Funnel™

2026-04-01

medium
General hospital and personal use Sterility
Quick summary

Quick summary: Flamingo Funnel™. Reason: Sterility. Category: General hospital and personal use. Date: 2026-04-01.

Recommended action

Recommended action: compare your product with the official notice and follow the instructions published by the official source.

Recall details

Quick summary: Flamingo Funnel™. Reason: Sterility. Category: General hospital and personal use. Date: 2026-04-01.

Official details

Affected product details

Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.

BrandNascent Surgical D/B/A SurgiSmoke Solutions
ProductFlamingo Funnel™
IssueMedical devices - Sterility
Recall classType II
IdentificationRA-81823

Affected products

ProductFlamingo Funnel™
Lot8264501
ModelSQ-20012-01
Serial number8264501
DescriptionFlamingo Funnel™
ProductFlamingo Funnel™
Lot8264601
ModelSQ-20012-01
Serial number8264601
DescriptionFlamingo Funnel™
ProductFlamingo Funnel™
Lot8264701
ModelSQ-20012-01
Serial number8264701
DescriptionFlamingo Funnel™

Issue

Packaging defect: A non-gamma-compatible drape degraded under heat, causing pinholes and possible loss of sterility. Users should immediately stop using affected lots, dispose of them per national and local regulations.

What you should do

  • Customers will be contacted with instructions to follow.
Additional official information
AudienceHealthcare
CategoryHealth products - Medical devices - General hospital and personal use
CompaniesNascent Surgical D/B/A SurgiSmoke Solutions 2300 Myrtle Ave, Suite 200, St. Paul, Minnesota, United States, 55114
Recall date2026-03-16
Published byHealth Canada
Recall classType II
Alert / recall typeHealth product recall
Identification numberRA-81823
Background{}
CompaniesNascent Surgical D/B/A SurgiSmoke Solutions 2300 Myrtle Ave, Suite 200, St. Paul, Minnesota, United States, 55114
AudiencesHealthcare
Source updated2026-04-01

Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.

Starting...

Related recalls

Similar recalls and safety notices.

Loading related recalls...