Quick summary: F&P Airvo 3. Date: 2024-08-29.
F&P Airvo 3
2024-08-29
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: F&P Airvo 3. Date: 2024-08-29.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
This recall relates to specific versions of software of the Airvo 3, and how the device responds when flow alignment alarm 3.2.2 occurs. When this happens in Airvo 3 devices set up with high pressure oxygen (HPO) and running software version 1.5.1 or earlier, the device will deliver room air only. If this happens, a patient may experience oxygen desaturation that could lead to hypoxia. Software version 1.5.2 updates the software algorithm and ensures the target therapy continues in the event that this alarm occurs. Recall Start Date: August 8, 2024
What you should do
- Customer will be contacted to arrange for software update.
Additional official information
Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.
Frequently asked questions
Answers based on the official recall notice.