Quick summary: Exoplan. Date: 2023-05-25.
Exoplan
2023-05-25
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Exoplan. Date: 2023-05-25.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
During an update of data used in a library shipped with Exoplan or made available on the download portal of Exocad, it was found that the use of a specific combination of an implant, sleeve and drill in a surgical kit can result in a hole being drilled for the implant 2mm too deep. In case an implant 260-340-256 of Bicon with a 4mm diameter and 6mm length is used in a fully-guided surgery procedure using a 4mm diameter Bicon sleeve it should only be possible to select a drill with 6mm length. But erroneously uses an 8mm length drill (instead of 6mm length drill) from the surgical kit. Recall start date: April 24, 2023
What you should do
- Contact the manufacturer if you require additional information.
Additional official information
Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.
Frequently asked questions
Answers based on the official recall notice.
Product images
Official product images from Health Canada.
Starting...
Explore more
Related recalls
Similar recalls and safety notices.