Evolve Triad Plate Cutter
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Evolve Triad Plate Cutter.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
Stryker has identified a nonconformance relating to the Evolve Triad Plate Cutter provided in the Evolve Triad Instrument Kit (4951kit1). Specifically, the plate cutter is unable to meet sterilization requirements when it is sterilized in a double wrapped configuration according to the option 1 steam sterilization parameters detailed in Instructions for Use (IFU150824). While the surgical technique does not require use of the plate cutter, it is provided in the instrument kit so the coronoid plate can be trimmed when needed. Recall start date: June 3, 2022
What you should do
- Contact the manufacturer if you require additional information.
Additional official information
Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.