Quick summary: Evis Exera III Video System - Gastrointestinal Videoscope. Date: 2024-11-01.
Evis Exera III Video System - Gastrointestinal Videoscope
2024-11-01
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Evis Exera III Video System - Gastrointestinal Videoscope. Date: 2024-11-01.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
Olympus previously issued revised cleaning instructions for the GIF-1TH190 on October 11, 2023, to issue new updated channel drying steps. Olympus conducted user evaluation, through human factors validation, and determined further clarification is required to improve the instructions for easier understanding of each drying step. As a result, Olympus is issuing a revised IFU addendum that informs customers that the channel drying time has been extended by 12 minutes to purge water from the channels. This is accomplished by drying each channel for 3 minutes, in the sequence described in the addendum. Ensure all reprocessing personnel are completely knowledgeable and thoroughly trained on the new water purge requirements. Olympus has received 16 reportable complaints related to this issue. Recall Start Date: October 18, 2024
What you should do
- Contact the manufacturer if you require additional information.
Additional official information
Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.
Frequently asked questions
Answers based on the official recall notice.
Product images
Official product images from Health Canada.
Starting...
Explore more
Related recalls
Similar recalls and safety notices.