Quick summary: EsoFLIP Dilation Catheter 30mm. Date: 2024-05-03.
EsoFLIP Dilation Catheter 30mm
2024-05-03
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: EsoFLIP Dilation Catheter 30mm. Date: 2024-05-03.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
Medtronic has received complaints of ES-330 catheters measuring diameters that are significantly (more than 50%) lower than actual. The failure mode may be reported as "failed precheck", "incorrect distensibility reading", "no distensibility value displayed", "adverse event without identified device or use problem", or "diameter incorrect." Medtronic Lafayette returned product analysis lab testing of unused, returned ES-330 catheters showed saline conductivity that is approximately 10 times lower than specification. The ES-330 saline conductivity specification is 2.25 mS/cm. The saline conductivity of multiple unused, returned ES-330 catheters had saline conductivity at or near 0.25 mS/cm. Recall start date: April 25, 2024
What you should do
- Contact the manufacturer if you require additional information.