Quick summary: Epoc Bgem Bun Test Card [25pk]. Date: 2024-10-04.
Epoc Bgem Bun Test Card [25pk]
2024-10-04
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Epoc Bgem Bun Test Card [25pk]. Date: 2024-10-04.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
Siemens Healthcare Diagnostics Inc. has confirmed that a negative sodium bias may exist in some card lots approximately 2 months prior to expiration. The observed maximum negative bias from internal testing is -10 mmol/l. No positive bias has been observed. Quality Control (QC) fluid may or may not detect the issue depending on the bias magnitude and the frequency of QC testing. Please note the card lots identified are now expired. Recall start date: September 25, 2024
What you should do
- Contact the manufacturer if you require additional information.
Additional official information
Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.
Frequently asked questions
Answers based on the official recall notice.
Product images
Official product images from Health Canada.
Starting...
Explore more
Related recalls
Similar recalls and safety notices.