Quick summary: Epiphany Cardio Server. Date: 2023-09-07.
Epiphany Cardio Server
2023-09-07
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Epiphany Cardio Server. Date: 2023-09-07.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
Epiphany, now a Baxter Healthcare company, is issuing an urgent medical device correction for the Cardio Server E3 ECG Management System Software due to a report of the caliper tool becoming non-responsive, preventing the caliper function from being used temporarily on software versions v6.1.x, v6.2.x, and v7.0.x of the E3 user interface. If this occurs, the calipers will appear on the screen but are not adjustable, and using any of the caliper controls results in no action. This issue does not prevent the user from changing measurement values before confirming a study. A software patch has been released to address this issue and our cardio server support team will contact you soon to arrange for installation. This issue does not affect the classic cardio server user interface. Recall Start Date: August 31, 2023
What you should do
- Contact the manufacturer if you require additional information.
Additional official information
Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.
Frequently asked questions
Answers based on the official recall notice.
Product images
Official product images from Health Canada.